How Do Universal Testing Machines Use Load Cells?
Universal testing machines (UTMs) are essential for material qualification, structural research, and quality assurance. Interface load cells sit directly in series with the moving crosshead and the upper specimen grip of a UTM. A machine controller reads the load cell continuously within a closed feedback loop to regulate crosshead travel, maintain constant loading rates, or hold targeted force thresholds during stress relaxation evaluations. The transducer measures the specimen resistance as it deforms. The data acquisition system pairs this instantaneous force data with displacement data to measure elasticity, yield point, and ultimate tensile strength.

Natural Frequency 101
Interface Natural Frequency 101 defines the inherent rate at which a sensor oscillates when allowed to vibrate freely. Learn how natural frequency and deflection affect dynamic force measurement, how to avoid resonance, and engineering tips for optimizing load cell accuracy. Natural frequency serves as the definitive boundary for the bandwidth of a force measurement system and for structural integrity. Accurate data acquisition from load cells relies on maintaining a clear separation between the application's operational frequency and the resonant frequency of the physical test loop.