Tag Archive for: voltage

Detailing Sensor Rated Output Millivolts Per Volt
Interface sensor specifications state rated output, expressed in millivolts per volt (mV/V), as the electrical output of a load cell, multi-axis sensor, or reaction torque transducer in mV/V of excitation at the rated load. It essentially expresses a load cell's sensitivity. RO represents the change in output voltage in millivolts of the load cell relative to a change in the excitation voltage in volts applied to it.

Excitation Voltage 101
An excitation voltage is needed to power the load cell and ensure the output signal is accurate. Load cell excitation is necessary to ensure the accuracy and reliability of load cell measurements. Interface provides electrical performance data on all specifications represented as VDC MAX, when applicable. The data for excitation voltage is listed under the electrical section of a transducer model’s specification datasheet, along with other factors, including rated output, bridge resistance, and zero balance.

Demystifying Specifications Webinar
Interface's technical force measurement webinar Demystifying Specifications details descriptions, terms, values and parameters found in product datasheets for load cells, torque transducers, instrumentation and specialty products. Learn from our experts what specifications need critical review, recommendations based on product categories, and the insider point of view on what is most important in terms of specifications for different use cases and tests.