Tag Archive for: SSMH

Tips for Properly Installing S-Type Load Cells
S-type load cell installation success hinges on protecting the sensor's gauged area from extraneous mechanical forces during setup. Interface provides installation tips, a setup video, and instructions. By ensuring that all tightening torque is applied to a single end, maintaining a minimum thread-depth engagement of one diameter, locking the threads to a shoulder or jam nut, and using high-grade hardware for loads exceeding the sensor’s capacity, you eliminate the leading causes of failure. Proper installation protects the load cell's calibration integrity, longevity, and safety across applications.

Mastering Fatigue Testing with Force Measurement Webinar Recap
Interface recorded webinar, Mastering Fatigue Testing with Force Measurement, delves into the intricacies of fatigue testing, fatigue-rated sensors, engineering tips, best practices, and frequently asked questions. The event details fatigue principles, standard testing techniques, and sensor-related equipment for practical fatigue analysis using force measurement data. Interface force measurement experts answer critical questions and provide engineering tips for fatigue testing using load cells.