Non-Repeatability 101
Interface 101 defines non-repeatability as the maximum difference between sensor output readings when the exact same load is applied repeatedly under identical loading and environmental conditions. Using load cells, understanding non-repeatability, why it governs a test's practical resolution limit, and how to isolate it from external test-rig anomalies is essential to designing dependable test protocols and accurate analysis.

Resistance Output Versus Rated Output
Rated Output (RO) is the dynamic electrical signal that directly represents the applied mechanical force. Resistance Output is a static electrical characteristic of the load cell's internal circuitry. When you are performing force measurements, you are primarily interested in the Rated Output. This value, in conjunction with your excitation voltage, enables you to accurately convert the measured millivolt signal into engineering units of force in lbf or Newtons. Discover how load cells work at the ultimate force measurement resource hub, Interface ForceEDU.

Detailing Sensor Rated Output Millivolts Per Volt
Interface sensor specifications state rated output, expressed in millivolts per volt (mV/V), as the electrical output of a load cell, multi-axis sensor, or reaction torque transducer in mV/V of excitation at the rated load. It essentially expresses a load cell's sensitivity. RO represents the change in output voltage in millivolts of the load cell relative to a change in the excitation voltage in volts applied to it.